- Built specifically for the Test described in IEC 60950, Annex N, Table N.1, Line 1 and IEC 62368-1, Annex D, Table D.1, Line 1.
- Charges the bulk capacitor to the required 2.5kV or 7kV (Australian Deviation), using the front panel knob or optionally via USB, see TMM-PF option.
- Discharge of the 6838V output pulse is controlled by the front panel TEST button or optionally via USB, see TMM-PF option.
- Meter reads the bulk capacitor voltage per the requirements of IEC 60950.
- Polarity can be changed by the front panel POLARITY button or optionally via USB, see TMM-PF option.
- Cables and NIST Calibration Certificate Included, A2LA Calibration optional.
- One year warranty and unlimited telephone support.
>
MegaPulse 10700-7PF, bulk cap voltages of 2000-7000V, 10x700uSec Voltage Waveform, 40 ohm 1 chnl. Designed specifically for testing to IEC 60950 and IEC 62368-1, including the Australian Deviation. Includes self-regulating high-voltage power supply for more stable voltage control.
Optional TestMinder PF computer control and 1000:1 Oscilloscope output (reference only). Option IER equips the tester with a switch compliant with IEC 60065, Fig. 7B.
In accordance with the requirements of IEC 60950, IEC 62368-1 and ITU-T Standards, the meter reads the bulk capacitor voltage.
Calibrated at bulk cap voltages of 1000-7000V to reflect the test points of IEC 60950, IEC 62368-1 and the Australian Deviation. ISO 17025 calibration available.